3He System D - Cell 12, 35 T
Four probes are available for use in both inserts of the 3He System D, located in cell 12 with a maximum field of 35T.
3He Insert
This system is a top-loading into liquid sorption pumped 3He cryostat. The temperature range of the insert is .3K to 80K. The available sample space I.D. is 17mm. Experiments performed include, but are not limited to, transport, magnetization, surface conductivity and optics with fibers. Rotating and fixed probes are available for use.
VTI Insert
This system is also a top-loading cryostat with a temperature range of 1.5 to 325K. The sample is loaded into dynamic vapor/liquid. The available sample space is 17mm and uses the same probes as the 3He insert.
For sample space dimensions on this system, see our Sample Space Dimensions table (PDF).
Click on the links below for PDF files with more details on this system.
Other Cryostats, Dewars, Probes and Sample Holders
For more information, contact DC Instrumentation Scott Maier.